Protective Cloth Synthetic Blood Penetration Tester

Adhering to American, British or European norms for visual assessment of color under standard light conditions. Visual assessment area two feet.

Light Sources:
D65, TL84, CWF – 2 Nos each
UVB – 1 No each.
Filament Lamps – 4 Nos.

Adherent Standards:
ISO 3664 / ASTM D1729 / BS950 / DIN6173

About the instrument:

Vital Medical protective clothing Synthetic Blood Penetration Tester is used to measure the penetration of synthetic blood and other liquid-derived liquid injection, to evaluate the safety of Medical protective clothing.

This instrument is suitable for the penetration test of synthetic blood spillage in medical protective clothing. This instrument is widely used in medical inspection department, safety inspection department and scientific research unit experiments, etc.

Test Principle:

The resistance of a protective clothing material to penetration by blood and body fluids is determined by subjecting the material to synthetic blood as a body fluid stimulant for a specified time and pressure sequence and observing if visible penetration of the liquid occurs. In the penetration test apparatus, the clothing material acts as a partition separating the body fluid simulant from the viewing side of the test cell. Any evidence of synthetic blood penetration constitutes failure.

Standards:

  • BS ISO 16603-2004:Clothing for protection against contact with blood and body fluids —Determination of the resistance of protective clothing materials to penetration by blood and body fluids —Test method using synthetic blood
  • ASTMF1670-2008:Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Synthetic Blood
  • ASTM F1670/1671:Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Synthetic Blood

Specifications

Operating System Touch screen Operating System
Test environment Temperature (21 + 5) ℃, Relative humidity (85±10)%
Pressure 1.75kPa   3.5kPa  7kPa  14kPa  20kPa
Power AC220V, 50Hz
Specimen Size 75mm×75mm
Pressure Area 28.27cm2
Retardation Net Stainless steel; open space ≥ 50%; bending ≤ 5mm under 14kpa;
Clamp torque 13.5N·m